Control Kit for the Calibration of RH Sensors

Control Kit for the Calibration of RH Sensors

The control kit consists of:

  • 5 vials of the same humidity value
  • 7 paper test strips
  • 1 control chamber for the version HKCxxCxx, without control chamber for version HKCxxS00
  • Box dimensions: 100 x 140 x 40 mm / 3.93 x 5.51 x 1.57" 0.1 kg
  • Weight: 1.0 kg / 2.2 lb

Accuracy:

  • An accuracy of ±3% RH can be achieved with strict adherence to test procedure.

Operating Conditions:

  • Reference temperature: 23°C ±1°C / 73°F ±2°F
  • Environmental humidity stabilized ±10%
  • Set up conditions:
    • Temperature between 0 and 50°C / 32 and 122°F with correction to be applied according to the instruction manual
    • Ambient RH between 40 and 60% RH
    • The instrumentation should be set up and the control kit must be kept in the same environment at least 10 hours before the procedure is started
  • Shelf life of the solutions:
    • In vial: unlimited
    • in the control chamber: 2 hours